The primary technical focus in the cardiac catheterization
laboratory is the generation, recording, and display
of high-quality x-ray images during diagnostic and interventional
catheterization procedures. With greater numbers of
increasingly complex interventional procedures being
performed, acquisition of fluoroscopic and cineangiographic
images of the highest quality remains crucial for the
optimal performance of the catheterization laboratory.
The ongoing trend toward more complex interventional
procedures results in greater exposure to radiation
for the patient and laboratory staff (93).
The longer procedure times associated with these procedures
also place greater demands on the x-ray generator and
tube than may have been the case previously. During
the last half-decade, the prominent role of 35-mm cine
film as the recording and archiving medium has been
challenged, and cine-less operation has become accepted
as routine practice in many laboratories (94).
The movement toward digital technologies in the catheterization
laboratory and throughout the hospital and community
continues to change the interaction between the laboratory
and the outside world. The rapid evolution in information
technology in turn changes users' expectations regarding
the accessibility of medical image data (95).
These significant technical changes require that the
basic requirements in the catheterization laboratory
be revisited with some frequency to ensure that the
technical needs of the laboratory are being met appropriately.
A. Radiographic Equipment
The conflict between acquisition of high-quality angiographic
images and limiting x-ray exposure of patients and staff
has always been difficult to resolve satisfactorily.
Cardiac angiography, with its simultaneous requirements
for high acquisition rates and the need to visualize
very small objects, places some of the most severe demands
on x-ray generating equipment. Although there has always
been optimism that improvements in technology would
help resolve the conflict, the increasing clinical requirements
instead have led to yet greater demands on the equipment.
The wide acceptance of digital angiographic systems,
which have advantages in a variety of procedures, has
led to new challenges and concerns about issues of data
rates, amounts, and storage (96,97).
High-quality video display has become standard in the
laboratory, and the use of pulsed-progressive fluoroscopy
is assumed with any currently available equipment (98).
B. Generators
The rising proportion of interventional procedures performed
in the catheterization laboratory increases the demands
placed on the x-ray angiographic equipment, including
the x-ray generator. These requirements typically include
a high-frequency generator with outputs of 80 to 100
kW at x-ray pulse rates of 30 pulses per second (60
pulses per second for pediatric applications). The ability
to perform pulsed fluoroscopy and angiography with short
exposure timessufficient to avoid motion blur
of objects moving at high speeds but still able to maintain
a high degree of contrasthas become a minimum
requirement in angiography. Modern equipment designed
for the catheterization laboratory should also have
automatic exposure control (AEC), which provides the
optimal combination of x-ray tube voltage, current,
and exposure time most suited for visualization of rapidly
moving coronary arteries with adequate contrast. Many
laboratories choose to use the high-level control (HLC)
fluoroscopic technique, which can produce exposure rates
beyond the standard regulatory limit of 10 roentgens
per minute (R/min) but less than that used during cineangiography.
There are broader implications for this higher exposure
mode related to the resulting exposure to the patient
and staff, but if the capability is deemed a requirement
for a particular laboratory, a generator with such capability
will be required regardless.
C. X-Ray Tubes
Along with the high-output generator, the x-ray tube
used in cardiac catheterization laboratoriesespecially
for long, complex interventional proceduresmust
meet the most demanding technical requirements (99).
To visualize the smallest coronary arteries and complex
variations in lumen geometry, focal spots of 0.6 to
0.8 mm are necessary. To acquire multiple angiographic
sequences lasting up to 10 to 30 s at rates of 30 frames
per second (and higher), the tube must be able to absorb
large amounts of energy and dissipate the resulting
heat quickly to avoid delays between acquisitionsand
serious damage to the x-ray tube. Similarly, the extensive
fluoroscopic times required for interventional procedures
will be limited by the heat dissipation characteristics
of the x-ray tube. Heat storage capacities >1 million
heat units (HU) and even approaching 3 million HUs have
been found invaluable in the catheterization laboratory.
A tube with greater heat storage capacity provides several
significant advantages in the modern clinical environment:
(1) it reduces delays during clinical procedures if
lengthy fluoroscopic and angiographic exposures result
in the heating of the tube to its maximum capabilities;
(2) it allows for penetration of larger patients (or
at steeper angulations) without resorting to higher
x-ray tube energies and the resulting reduction in vessel
(and device) contrast and increase in image noise; (3)
it provides the option for use of filtering materials
that can either reduce patient skin exposure, improve
image noise characteristics, or both.
D. Image Intensifiers
The performance characteristics of image intensifiers
used in the catheterization laboratory have improved
continuously over the years. Higher conversion factorsgreater
efficiency in light output as a function of x-ray input
exposurealong with improved contrast ratios and
a reduction in geometric distortion have led to improved
image quality. As a result, image intensifier performancein
dedicated cardiac catheterization systemshas been
optimized for the task of imaging the heart and coronary
arteries at the x-ray exposure levels used in that application.
In general terms, this translates to a high-contrast
spatial resolution >3.0 line pairs/mm at the entrance
plane of the image intensifier and acceptable signal-to-noise
quality at cineangiographic entrance exposures of 20
to 25 microR/frame in the typical magnification mode
used for coronary angiography (100).
On the other hand, systems optimized for other imaging
tasks, such as peripheral and vascular angiography,
cannot at the same time deliver the required performance
in the heart, and users should be aware that there will
be degradation in performance for the task of imaging
the heart and coronary arteries. The different x-ray
tube design, e.g., target size and angle, required for
covering a field of view corresponding to a 40-cm image
intensifier will not deliver the same results at the
same x-ray exposure for the 12- to 20-cm field of view
customary in cardiac imaging. Similarly, the light-gathering
characteristics of the larger image intensifier may
compromise performance for cardiac imaging at the smaller
fields of view when only a fraction of the intensifier
input surface is used. In turn, there are implications
for the electrostatic optics within the intensifier
as well as for the light-gathering optics at the output
surface of the intensifier; in general, these effects
will degrade the image quality relative to an intensifier
specifically designed for the cardiac application. In
summary, the effects that can occur in such a system
are degradation in spatial resolution, increased image
noise in fluoroscopy and angiography, increased x-ray
exposure rates, and delays due to exceeding heat capacity
for x-ray tubes.
E. Developments in X-Ray Detectors
As mentioned above, the image intensifier is a vital
component for x-ray angiographic imaging in the cardiac
catheterization laboratory, because no better alternatives
exist (to date) that can convert the x-ray intensity
information exiting the patient into usable, visible
light information. This visible light is converted to
an electrical video signal and in turn to a stream of
numbers for subsequent processing and storage. Recent
developments in digital x-ray detector technology demonstrate
significant potential for application in cardiac fluoroscopic
and angiographic applications in the future. The first
commercial systems have only recently been introduced,
and it is expected that they will become more common;
readers should thus be aware of how these systems differ
from those to which they have been accustomed. The most
significant difference between the image intensifier-based
imaging chain and those based on digital detector technologies
is that the entire image intensifier tube will disappear,
along with the video camera used to convert the output
light signal to an electronic voltage. The promise is
that this will deliver significant improvements in spatial
resolution because it will not be limited by "blurring"
processes inherent to the image intensifier systems,
greater dynamic range and contrast resolution, and because
the new detectors are much simpler mechanically, with
less likelihood of failure and gradual degradation in
performance (101).
The candidate detectors that will most likely be appearing
in cardiac catheterization laboratories are digital
flat-panel detector systems that use a compact
box in place of the image intensifier tube to convert
the incident x-ray signal directly, essentially to an
array of discrete electrical signals that are read individually
and processed, displayed, and stored for further processing
and review. These types of detectors are characterized
as direct digital because the intensity values are generated
as an array of digital values (ones and zeros) at the
detector without the need for additional conversion
processes that can add noise or degrade performance
(102).
These detectors have been made possible by improvements
and cost reductions in the manufacture of flat-panel
displays used in the computer industry (e.g., laptop
computer monitors). Essentially, these active matrix
arrays are combined with an x-ray-sensitive layer that
converts the x-ray signal to light incident on an array
of light-sensitive cells, anywhere from 1000X1000 to
2000X2000 cell arrays. Clinical testing of the first
of these devices has begun, and these detectors are
now becoming available as product options.
F. Video Components
1. Video Cameras
Along with image intensifiers, high-quality video cameras
have been an assumed component of modern cardiac angiographic
systems for generation of high-quality images during
fluoroscopy and to provide the analog signal source
for the conversion process used in all current-generation
digital angiographic systems. The traditional "pickup
tube" camera, based on a scanning electron beam to read
off the spatially varying electrical signal produced
by the light output of the image intensifier, is well
understood and has been described in detail in the previous
guidelines (5).
Modern systems use cameras that operate in "standard"
resolution mode (525 lines per video frame) as well
as "high" resolution mode (1023 or 1049 lines) for both
fluoroscopic and angiographic applications. An important
characteristic that should be carefully assessed in
these systems is the introduction of additional electronic
noise to the image in the higher-resolution modes requiring
higher bandwidth electronics. Another aspect related
to the discussion of "dual-use" systems above relating
to image intensifier performance is the fact that video
systems designed for slow frame rate angiographic applications
(typically at higher x-ray entrance exposures) may demonstrate
degraded temporal performance (i.e., blurring) when
operated at the faster acquisition rates required for
cardiac imaging.
Although the pickup tube video camera has long been
the workhorse of cardiac angiographic systems, a relatively
recent development has been the increasing availability
of video cameras based on solid-state image sensors
(e.g., the charge-coupled device or CCD cameras) (103).
CCD sensors consist of an array of discrete elements
(typically 1024 X 1024) that store the light information
from the image intensifier output until they are read
by the camera electronics. Among the advantages of the
CCD camera are simpler design, resulting in smaller
size; improved dynamic range; improved spatial resolution;
absence of temporal lag; prolonged life; lower cost
and simplified maintenance requirements; and lack of
sensitivity to magnetic fields. (Note that the flat-panel
detectors described above also share a number of these
advantages.) The advantages listed above have been extensively
demonstrated and, in general, CCD cameras operating
at the same pixel resolution do offer better image quality
than video tube cameras. It should be understood, however,
that the use of CCDs in the detector chain does not
itself make this a direct digital detector. In most
applications, the output of the CCD camera is a time-varying
analog voltage signal that must be digitized before
use as with pickup tube cameras. One issue that should
be considered is the increased amount of data that can
be generated by the larger matrix sensors.
G. Digital Angiography Issues
Among the most significant changes in practice in the
cardiac catheterization laboratory in recent years has
been the consistent move away from 35-mm cine film as
the standard recording medium. In many laboratories
in the United States as well as abroad, cine film is
no longer used in any function of the catheterization
laboratory. This has advantages that have been well
documented (103,104),
but there are also a number of changes in the technical
requirements for a laboratory and considerations that
must be made in equipment purchase and use.
H. Effects on X-Ray Requirements
For many years, the promise of digital angiographic
recording has been accompanied by the promise of reductions
in x-ray exposure (105).
This has in turn led to confusion when this reduction
has not materialized as digital angiographic systems
are implemented. The image quality in cardiac catheterization
specifically at the exposure per frame that is customarily
found in this application is primarily a function of
the contrast signal and the image noise that are found
acceptable for the application. With state-of-the-art
equipment, the primary source of image noise resides
in the image intensifier, and this may be objectionable.
The light output of image intensifiers designed for
cardiac catheterization is more than enough at standard
entrance exposures for both cine film and digital recording.
Elimination of the cine camera does not by itself mean
that more light is available for the video chain; there
are already adequate amounts of light to obtain good
video signals. However, the noise is directly related
to the statistical properties of the relatively low
x-ray beam flux incident to the image intensifier. It
is true that a digital-only angiographic system allows
more flexibility with the light-limiting apertures in
the video chain because one is not limited by the requirements
of delivering adequate light to the cine camera. But
the use of such apertures as a dose-adjustment method
is not routine. More practically, the elimination of
cine film does allow use of reduced frame rate acquisition
with the proportional reduction in x-ray exposure (19).
In the end, however, x-ray exposure per frame is affected
more by other factors, such as those described above.
I. Digital Acquisition Requirements
As discussed above, at the time of this writing, essentially
every digital angiographic system requires a conversion
from an analog signal produced by a video camera to
a string of numbers stored and processed for display
and analysis. The spatial, contrast, and temporal resolution
requirements for cardiac catheterization are well understood
and have been met for the most part with digital systems
operating at matrix sizes of 512 X 512 pixels, bit depths
of 8 bits (corresponding to 256 intensity values), and
acquisition rates of 30 frames per second. With the
ongoing reduction in the cost of digital hardware, it
has been possible to deliver improved performance as
well. Newer digital systems can record to larger image
matrices (1024 X 1024 is common) and greater bit depths,
with 10- and 12-bit images becoming available. With
pickup tube cameras, higher spatial resolution is a
function of the camera scanning rate along with the
image intensifier magnification mode (and geometric
magnification). With CCD cameras, the size of the element
on the camera is necessarily fixed, but improvements
in resolution are still possible through the use of
higher magnification modes, which map a fixed-size element
to a smaller object in the patient. Similar factors
apply as well to digital detector technologies, but
in that case, the resolution cannot be improved through
selection of a higher magnification mode. In a field
of view of 15 cmtypical for coronary angiographya
512 matrix results in a limiting resolution of 0.20
to 0.25 mm (in the plane of the imaged object), corresponding
to a resolution of 2.0 to 2.5 lp/millimeter. While this
has been found in general to be adequate for coronary
imaging, it is less than the theoretical resolution
of cine film. Accordingly, many vendors are offering
higher resolution systems. In considering these systems,
however, it is important to ensure that the contrast
resolution is adequate as well. For instance, if there
is insufficient contrast from small objects due to other
factors in the imaging chainx-ray energy, image
intensifier or video issues, or radiation scatter
the improved spatial resolution will be of limited advantage.
The issue of much greater data storage requirements
must be considered as well.
J. Digital Storage and Display
In most digital angiographic systems in use today, there
is limited storage capacity on the system itself, usually
only enough for one to several days worth of procedures.
It is therefore necessary to have a mechanism for medium-
and long-term storage. The development of the DICOM
standard for cardiac angiography accelerated the use
of digital storage media by ensuring that there was
a well-understood method for exchanging digitally recorded
procedures between laboratories and systems (106,107).
Many laboratories purchased the capability for storing
exams on CD-ROMs and have pursued that as a long-term
storage medium. Other laboratories have implemented
automated storage libraries, which provide access to
many months or years of procedures without the need
for manual intervention for retrieval and display. The
specific archival systems used by a laboratory should
be selected on the basis of clinical and financial considerations
(96,97).
Whatever approach is taken, every laboratory should
ensure that the version of data retrieved from the archive
at some later date is identical to the version used
for postprocedure diagnosis and decision making. In
most laboratories the digital image data are typically
not reintroduced into the digital angiographic system
for later review, but rather are reviewed on a separate
digital image review workstation. This workstation can
be supplied by either the original x-ray equipment vendor
or increasingly from a variety of third-party manufacturers.
The quality and cost of such workstations can vary greatly.
A laboratory should ensure that review performance from
media or over a network is adequate for subsequent clinical
assessment. Among the factors that must be considered
are display rates equivalent to the original acquisition
rate, full image display resolution, image processing
and enhancement, and sufficient exam storage capacity
to avoid the need for delay in retrieval of older exams.
It should be noted that both the technology and cost
of imaging workstations are changing rapidly, and a
laboratory should anticipate future needs at the time
of equipment purchase.
K. Image Formats and Standards: The DICOM Standard
The acceptance of the DICOM standard for cardiac angiography
provided assurance that a version of an angiographic
exam exchanged with another laboratory or reviewed at
a later date could be identical to that reviewed in
the laboratory during and immediately after a catheterization
procedure. It is unfortunately not always the case that
every vendor claiming to subscribe to the standard delivers
equipment that does in fact meet the functionality stated
above. The DICOM standard does include a conformance
mechanism by which a vendor is required to list the
capabilities and supported features of the product,
but the standard remains relatively new, and unfortunately,
a laboratory cannot rely solely on such claims. Any
archiving, exchange, or review system should be carefully
assessed to determine whether the exam data stored for
local archiving or written to media for exchange do
in fact provide the same quality of diagnostic information
achieved in the original.
One factor that can ensure such equivalence is the
use of a direct digital interface between the angiographic
acquisition system and the archiving or review system.
Due to the relatively recent introduction of the standard
and the fact that there is a large installed base of
equipment of varying age, this digital interface is
achieved through a range of sometimes complicated approaches
or, for some of the oldest equipment, cannot be achieved
at all.
Newer digital angiography equipment is available with
a DICOM network interface, meaning that the exam information
leaving the system is already formatted according to
the DICOM standard. This has 2 advantages: (1) data
equivalence is assured and (2) any receiving system
that supports this interface can be used for storage
and review (i.e., a laboratory has more choices). An
alternative approach typically used with older equipment
is a digital interface, which requires an additional
step to format exam data to the DICOM standard. This
approach can work as well, but it requires that the
developer of the interface, usually the manufacturer
of the x-ray system, make the interface specification
available to other vendors. Otherwise, only equipment
from the original vendor will support the interface.
Again, it should be noted that either type of interface
will usually work, but it should be understood from
the onset which type is being provided.
Another alternative that has also been implemented,
especially with older equipment, is an analog capture
interface rather than a digital interface. In this approach,
the analog video signal is captured and digitized somewhere
in the acquisition or display chain, and this second
digital version is then formatted according to the DICOM
standard and used for archiving, display, and exchange.
Laboratories should be cautioned that the image data
resulting from this approach is not strictly equivalent
to the information available at the time of the procedure.
With appropriate precautions, the quality can still
be quite high, but, in many cases there is significant
degradation in image quality. In the case of older equipment
for which no alternative exists, this approach does
provide a digital archive and exchange approach of nearly
equivalent image quality, but only if the parallel image
capture is performed to a specification close to that
incorporated within the original x-ray vendor's analog-to-digital
conversion. In some cases, irreversible data compression
is also used during the capture process to save costs
and improve performance, but this leads to artifacts
in the stored archival copy, which degrade image quality
and can affect visualization of the anatomy.
L. Digital Image Resolution
As discussed earlier, digital cardiac angiographic systems
have most often incorporated digital resolutions of
512 lines by 512 columns by 8 bits per sampleusually
together with acquisition rates of 30 frames per second.
The initial basic version of the DICOM standard for
digital cardiac angiography supported only this format
to provide at least 1 format that many vendors could
support. In recent years the number of cardiac angiographic
systems available at higher resolutions has increased,
and the issue of digital equivalence between the acquired
exam information and the stored and exchanged versions
has emerged. The issues include the clinical requirement
for the higher resolution and whether the permanently
stored copy should also be stored with the higher resolution.
In general, a matrix size of 512X512 has been deemed
acceptable for clinical applications despite the fact
that this corresponds to a minimum spatial resolution
on the order of 0.2 to 0.3 mm.
For some applications, such as quantitative coronary
angiography of complex stenoses, higher resolution (e.g.,
0.1 to 0.15 mm) is seen as optimal (100).
In that case, the 1024 X 1024 matrix size will deliver
improved resolution but at the cost of increased data
acquisition, storage, and transmission requirements.
Under these circumstances, the version of the exam stored
locally as well as used for exchange should accommodate
the higher resolution images on which the clinical diagnosis
was made. The DICOM standard does accommodate these
higher image matrices, but a laboratory should ensure
that the equipment purchased for this application does
indeed store all the acquired information and can write
exchange media in the larger format (assuming that the
receiving laboratory in turn has the ability to display
the larger format images). Similar concerns apply to
network transfer, which will require greater bandwidth
capacity for the larger amounts of data or, alternatively,
greater delays in transmission.
M. Data Compression
Despite the rapid improvements in digital storage and
transmission technology, the data requirements of digital
cardiac angiography remain among the most demanding
in medical imaging. As a result, some equipment vendors
and suppliers of imaging applications have sought to
reduce the amount of data through the use of mathematical
techniques or compression methods (108).
Some compression methods are completely reversible or
lossless, and in the end the recipient or user has precisely
the same equivalent data that were available initially.
In contrast, irreversible or lossy compression methods
reduce the amount of data, but the resulting image information
is not strictly identical to the original angiographic
information. The variation that can be detected visually
and the effect it may have on clinical assessments made
from the images depends on the type of compression method
and degree of reduction. Lossy compression methods can
result in much greater amounts of data reduction, but
the resulting images may contain detectable artifacts
that were not in the original image. The basic problem
with the use of irreversible compression methods is
that a user at a later time is not provided with the
same information used initially; this may be acceptable
in some applications, but none in which the original
information is no longer available if needed. The ACC
and the European Society of Cardiology have sponsored
a multicenter clinical study to assess the effects of
one of the more common compression methods (motion JPEG)
on critical diagnostic tasks. The results of that study
indicate that as the amount of compression is increased,
the ability to detect clinical features is impaired.
It is strongly suggested that laboratories avoid the
use of lossy compression methods for the permanent storage
of digital angiographic data (109).
N. Telemedicine Applications
Routine storage and availability of angiographic records
in a digital format makes possible a new class of clinical
applications, which fall under the general category
of telemedicinereferring to the electronic transmission
of clinical image data over large distances to support
clinical decision making at remote sites (110).
In addition to the digital format of the procedure data
in the acquiring laboratory, this requires an accepted
standard format for the image data that can be displayed
at the receiving site as well as a reliable digital
network link between the sending and receiving centers.
Such Wide Area Network (WAN) applications extend, in
simplest terms, the network from within a laboratory
or hospital to much greater distances. The network in
effect provides a user hundreds of miles away with the
ability to display and review the procedure image data
as if he or she were in the procedure laboratory.
In one example of this type of application, simple
"store and forward" transmission of a procedure from
a referring hospital to another medical center may be
performed, in effect replacing the mail or courier service
with electronic transfer. One advantage of this approach
is that the original copy of the exam record remains
in the acquiring laboratory. At the other end of the
spectrum, "expert" physicians at one center can participate
in and provide advisory support in real time for a procedure
being performed at another center.
Although the cost of digital network transmission is
being reduced and networks are being extended to more
locations, the bandwidth needed to transmit cardiac
catheterization examination results rapidly and completely
remains costly and/or relatively rare at this time.
As with many networking applications, users must choose
between speed and expense. As a result, some attempts
to provide telemedicine services for catheterization
procedures in a less costly manner have incorporated
data compression methods to reduce the size of the data
files being transmitted and, in turn, reduce the time
required to transmit the files. As with the issues raised
in the discussion above, this means that the data being
reviewed at the distant location may differ to a clinically
relevant degree from the data used in the initial review
and diagnosis. There may be applications for which this
is acceptable, but laboratories and hospitals should
not use such systems for an application that involves
decision making solely on the basis of compressed images,
which are degraded in diagnostic quality. Specifically,
systems developed for routine video conferencing over
standard telephone lines do not as a rule have adequate
image quality for transmission of clinical image data
acquired under the conditions required for cardiac angiography.
Similar concerns apply to transmission over the Internet
using standard networks. There are technical means available
to deliver diagnostic image quality of digital angiograms
in a timely fashion; laboratories considering a telemedicine
application should examine the technical specifications
of such systems carefully. In the near future, clarification
of privacy concerns for individual patients should allow
for discussions of patient cases without jeopardizing
patients' confidentiality.
O. Quantitative Measurement Methods
With increasing storage and easy access to catheterization
results in a digital format, the use of computerized
methods to measure coronary artery stenosis severity,
left ventricular function, regional wall motion, and
other techniques becomes much more feasible (111,112).
These methods provide objective, reproducible measures
for assessment of disease and physiological function
and are available for use in all laboratories rather
than being limited to multicenter clinical trials. Laboratories
should ensure that systems being used have been fully
validated in well documented in vitro and in vivo studies
and that technicians are familiar with the procedures
necessary to produce measurements that are reliable,
accurate, and reproducible. With the elimination of
cine film and its replacement with digital images that
are readily accessible, one of the greatest sources
of variability is no longer an issue with the elimination
of the need to convert cine film to a digital file.
Consistent procedures must be followed, and it is important
that the analysis be performed on the original image
data acquired at the time of the procedure. As mentioned
above, systems are available that store and exchange
versions of the images that differ significantly from
the original acquired data; in general, results on such
secondary capture images or, possibly, compressed images
will not be as reliable.
P. Further Developments in the DICOM Standard
Just as the DICOM standard for digital images made it
possible for catheterization laboratories to routinely
exchange high-quality digital images, other developments
are under way to enable similar standardized formats
for other types of information gathered during catheterization
procedures (113).
An outgrowth of the DICOM effort extends the standard
format to include physiological data: hemodynamic and
electrocardiographic waveforms and measurement results.
The primary focus of the effort was to facilitate a
"unit patient record" for the catheterization procedure
that includes all information necessary as a local record
and for interchange between laboratories and hospitals.
Although the standard is in a preliminary stage at this
time, newer versions of imaging and waveform recording
equipment will support the standard, and laboratories
that wish to incorporate such capabilities should obtain
assurance that all required equipment supports the standard.
In a related development, standard formats can ensure
that the results of the procedure data can be exported
electronically in a format compatible with national
data registries such as the ACC-NCDR (114).
Other developments include the integration of other
imaging modalities such as intravascular ultrasound,
as well as standard formats for clinical reports that
can be transmitted electronically in such a manner that
they can be displayed in other laboratories regardless
of which vendor's equipment was used to generate the
original procedure report.